posted on 2024-09-06, 05:50 authored by V.R. Mundembe
This paper presents an example of a typical calibration of a partially immersed Liquid-in-Glass (LiG) thermometer, by comparison to a standard precision digital thermometer, to obtain a correction to the LiG thermometer at a nominal temperature of 50° C. The standard way of evaluating and expressing the measurement uncertainty in typical corrections and uncertainty sources for the calibration has been used.

A research paper on scientific measurement.

History

Publisher

University of Zimbabwe (UZ) Publications

Citation

Mundembe, V.R. (2001) Uncertainty in the calibration of a partially immersed Liquid-in-Glass thermometer, Journal of Applied Science in Southern Africa, vol. 7, no.2, pp. 121-131. Harare: UZ Publications.